Health Products

Online DAT Amphetamines II for Cobas Integra

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
65423601
Manufacturer
Roche Diagnostics GMBH
Sandhoffer Strasse 116
Mannheim
68305
GERMANY
A. Online DAT Amphetamines II for Cobas Integra4512936190

Introduction

Recalled Products
  1. Online DAT Amphetamines II for Cobas Integra

Reason

Patient and control sample recovery may be decreased with the online dat amphetamines II reagent (Lot number 654236-01) when used on the Cobas Integra 400 and 800 Series Analysers. The affected lot shows reduced D-Amphetamine cross-reactivity for higner concentrated D-Amphetamine samples.