· Published by Health Canada · Official notice
This is duplicated text of a letter from ArjoHuntleigh Canada Inc. Contact the company for a copy of any references, attachments or enclosures.
Notice about Health Canada advisories
May 3, 2013
Dear Health Professional,
Please distribute this document to all personnel involved in the day to day usage of the Alenti Lift and Hygiene chair and post this NOTICE in your facility.
Subject: Alenti Lift and Hygiene Chair – Risk of Chair Tipping and Patient Falls
ArjoHuntleigh, in collaboration with Health Canada, is providing this notice to re-iterate safety information to prevent the risk of chair tipping and patient falls with the Alenti Lift and Hygiene chair. These incidents can be caused by operator use error and patient factors, and can lead to serious patient injury or death.
Between 2004 and 2007, ArjoHuntleigh conducted Field Safety Correction Actions to provide improved safety instructions and warnings to reduce the risk of Alenti chair tipping and patient falls. The incident rate has since declined; however, ArjoHuntleigh continues to receive several incident reports in Canada each year. To further reduce the number of incidents associated with the Alenti chair, ArjoHuntleigh would like to re-iterate the important safety information.
Important Safety Warnings – Risk Factors Associated with Chair Tipping and Patient Falls
The majority of chair tipping and patient fall incidents occur in the hygiene bathing room. Special attention should be taken to precisely follow the Instructions for Use regarding the bathing and patient drying procedures.
WARNING - The following precautions must be observed at all times when operating the Alenti. Failure to observe these precautions and failure to strictly follow the Instructions for Use can cause serious patient injury.
1) PATIENT ASSESSMENT
2) SAFETY BELT
3) INTERFERENCES
4) ELEVATION
5) SUPERVISION
6) REPOSITIONING
Important Information on Alenti Maintenance and Inspection
Incidents have occurred as a result of the Alenti wheels becoming detached during use. The Preventative Maintenance Schedule found in the Instructions for Use must be followed, with special attention to weekly inspection of the wheels to ensure that they are properly fixed and rolling and swiveling freely. Before each use, check that all parts are in place and DO NOT use the Alenti if any part is missing or damaged. Failure to maintain and inspect the device correctly can result in serious patient injury.
Important Information on the Alenti Operating Wall Chart
ArjoHuntleigh has revised the Alenti Operating Wall Chart. The changes in the Wall Chart include:
Confirming Receipt of this Alenti Notice to Hospitals
As confirmation that you have read and understood this Notice to Hospitals and have received the new and improved Alenti Operating Wall Chart, please complete the Customer Response Form online using the information provided on the enclosed postcard.
If you have any further questions or require assistance completing the Customer Response Form online, please contact ArjoHuntleigh at 1-855-312-4265 or via email at AHNA-RA@arjohuntleigh.com.
Reporting Adverse Incidents
Managing marketed health product-related adverse incidents depends on health care professionals and consumers reporting them. Reporting rates determined on the basis of spontaneously reported post-market adverse incidents are generally presumed to underestimate the risk associated with health product treatments. Any cases of Alenti chair tipping or patient falls or other incidents involving Alenti chairs should be reported to ArjoHuntleigh or Health Canada at the following addresses:
ArjoHuntleigh Canada Inc.
300-90 Matheson Blvd. West
Mississauga, Ontario L5R 3R3
Telephone: 1-800-665-4831
E-mail: info.canada@arjohuntleigh.com
To correct your mailing address or fax number, contact ArjoHuntleigh Canada Inc.
Any suspected adverse incident can also be reported to:
Health Products and Food Branch Inspectorate
HEALTH CANADA
Address Locator: 2003D
Ottawa, Ontario K1A 0K9
Telephone: The Inspectorate Hotline 1-800-267-9675
The Medical Devices Problem Report Form and Guidelines can be found on the Health Canada Web site.
For other medical device inquiries related to this communication, please contact Health Canada at:
Marketed Health Products Directorate
E-mail: mhpd_dpsc.public@hc-sc.gc.ca
Telephone: 613-954-6522
Fax: 613-952-7738
Regards,
original signed by
Chris Sloan
Quality Assurance Manager
ArjoHuntleigh Canada Inc.