Health Products

Exactrac Patient Positioning System

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Software Version 6.0.1 and 6.0.2
Manufacturer
BrainLab AG
Kapellenstr. 12,
Feldkirchen
85622
GERMANY
A. Exactrac Patient Positioning System49931

Introduction

Recalled products
  1. Exactrac Patient Positioning System

Reason

After performing a snap verification, that is a monoscopic intra-fraction X-Ray monitoring using only one X-Ray tube, in combination with implanted markers, the Exactrac 6.0 X-Ray monitoring tab might display an incorrect hint icon and resume text.