Becton Dickinson Vacutainer Plus Plastic Plasma Separator Tube (PST)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer 
Manufacturer
Becton Dickinson and Company
1 Becton Drive
Franklin Lakes
07417
New Jersey
UNITED STATES
A. Becton Dickinson Vacutainer Plus Plastic Plasma Separator Tube (PST)367960

Introduction

Recalled products
  1. Becton Dickinson Vacutainer Plus Plastic Plasma Separator Tube (PST)

Reason

A product defect named "lip-out" was detected for BD Vacutainer PST tubes. It consists of gaps of various sizes in the structure of the tube wall that is present at the lip of the tube. These gaps are covered by the stopper and hemogard closure, so the tube can be evacuated and used for blood collection. Customer complaints cite jamming of preanalytical automation (tube decappers) and leakage of blood in the specimen transport bag.