OCD has received several reports of condition codes that occurred during the calibration of Microwell Assays. Internal investigation determined that the Microwell Shuttle Weight on some Vitros systems may be uncoated and, therefore, out of our manufacturing specifications. If an uncoated Microwell Shuttle Weight is installed and calibration of Microwell assays is not performed, biased results could potentially occur. Internal testing determined that quality control results using an uncoated well weight showed, on average, a 10% shift from established means that was within the quality control ranges. The magnitude of the bias is not likely to impact patient safety.