Health Products

Total Protein 2 Reagent (Urine) for the Advia Chemistry Systems

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
149411B. Advia 1200 Chemistry Total Protein 2 System
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave.
Tarrytown
10591
New York
UNITED STATES
05000171
149411A. Advia Chemistry Total Protein 2 (Urine) System
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave.
Tarrytown
10591
New York
UNITED STATES
05000171

Introduction

Recalled products
  1. Advia Chemistry Total Protein 2 (Urine) System
  2. Advia 1200 Chemistry Total Protein 2 System

Reason

Siemens Healthcare Diagnostics is conducting a recall for the Total Protein 2 Reagent (Urine)(UPRO2) lot 149411 on the Advia Chemistry systems. Due to precipitation of the reagent indicator dye in this lot, an increase in failed calibrations and quality control failures has been observed. These failures will result in no patient result being reported with the affected kits.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime