Health Products

Advia Centaur and Centaur CP Systems - TSH3-Ultra Assay

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
All lotsB. Advia Centaur CP System- TSH3-Ultra (TSH3-UL) Assay
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave.
Tarrytown
10591
New York
UNITED STATES
06491072
All lotsA. Advia Centaur System - TSH3-Ultra (TSH3-UL) Assay
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave.
Tarrytown
10591
New York
UNITED STATES
06491072

Introduction

Recalled products
  1. Advia Centaur System - TSH3-Ultra (TSH3-UL) Assay
  2. Advia Centaur CP System- TSH3-Ultra (TSH3-UL) Assay

Reason

Siemens Healthcare Diagnostics is informing customers that a rare variant of TSH, identified in a small cluster of patients, is not detected by the Siemens Assays.  As part of an investigation of unexplained discordant results, samples were identified in which the monoclonal antibody used in the reagent failed to detect the TSH molecule. These individuals may have a previously unrecognized, functionally normal TSH variant. The observed rate of occurrence during a 30-month time period was 0.6 x 10-7.