Dimension Chemistry Systems May 27, 2013 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : General Public, Healthcare Professionals, Hospitals
Department : Health Canada
Hazard classification : Type II
Lot or serial number Companies Brand name Model or catalog number 2KD024 Manufacturer Siemens Healthcare Diagnostics Inc. 500 GBC Drive, Mailstop 514, PO Box 6101 Newark 19714-6101 Delaware UNITED STATES B. Dimension Xpand System-LMMB Mass Creatine Kinase MB Isoenzyme Flex Assay RC420 3AD051 Manufacturer Siemens Healthcare Diagnostics Inc. 500 GBC Drive, Mailstop 514, PO Box 6101 Newark 19714-6101 Delaware UNITED STATES C. Dimension EXL System-Mass Creatine Kinase MB Isoenzyme (MMB) Assay RC420 2GD053 Manufacturer Siemens Healthcare Diagnostics Inc. 500 GBC Drive, Mailstop 514, PO Box 6101 Newark 19714-6101 Delaware UNITED STATES A. Dimension Chemistry System-Mass CKMB RC420
Recalled products
Dimension Chemistry System-Mass CKMB
Dimension Xpand System-LMMB Mass Creatine Kinase MB Isoenzyme Flex Assay
Dimension EXL System-Mass Creatine Kinase MB Isoenzyme (MMB) Assay
Siemens has received customer complaints that the lyophilized cake is not fully dissolved after the stated time in the mmb calibrator instructions for use (IFU). The frequency of this occurrence is low but if it occurs, gel-like clumps may be observed.