Health Products

Immulite Systems - Rapid Thyroid Stimulating Hormone (TSH) Assays

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots
Manufacturer
Siemens Healthcare Diagnostics Products Ltd.
Glyn Rhonwy
Llanberis, Gwynedd
LL55 4EL
UNITED KINGDOM
F. Immulite 1000 System - Third Generation TSH AssayLKTS1
All lots
Manufacturer
Siemens Healthcare Diagnostics Products Ltd.
Glyn Rhonwy
Llanberis, Gwynedd
LL55 4EL
UNITED KINGDOM
D. Immulite System - Third Generation TSH AssayLKTS1
All lots
Manufacturer
Siemens Healthcare Diagnostics Products Ltd.
Glyn Rhonwy
Llanberis, Gwynedd
LL55 4EL
UNITED KINGDOM
E. Immulite 2000 System - Third Generation TSH AssayL2KTS2
All lots
Manufacturer
Siemens Healthcare Diagnostics Products Ltd.
Glyn Rhonwy
Llanberis, Gwynedd
LL55 4EL
UNITED KINGDOM
B. Immulite 2000 System - Rapid Thyroid Stimulating Hormone (TSH) AssayL2KRT2
All lots
Manufacturer
Siemens Healthcare Diagnostics Products Ltd.
Glyn Rhonwy
Llanberis, Gwynedd
LL55 4EL
UNITED KINGDOM
C. Immulite 1000 System - Rapid TSH AssayLKRT1
All lots
Manufacturer
Siemens Healthcare Diagnostics Products Ltd.
Glyn Rhonwy
Llanberis, Gwynedd
LL55 4EL
UNITED KINGDOM
A. Immulite System - Rapid TSH AssayLKRT1

Introduction

Recalled products
  1. Immulite System - Rapid TSH Assay
  2. Immulite 2000 System - Rapid Thyroid Stimulating Hormone (TSH) Assay
  3. Immulite 1000 System - Rapid TSH Assay
  4. Immulite System - Third Generation TSH Assay
  5. Immulite 2000 System - Third Generation TSH Assay
  6. Immulite 1000 System - Third Generation TSH Assay

Reason

Siemens Healthcare Diagnostics is informing customers that a rare variant of TSH, identified in a small cluster of patients, is not detected by the siemens assays. As part of an investigation of unexplained discordant results, samples were identified in which the monoclonal antibody used in the reagent failed to detect the TSH molecule. These individuals may have a previously unrecognized, functionally normal TSH variant. The observed rate of occurrence during a 30-month time period was 0.6 x 10-7.