Terumo Advanced Perfusion System 1 (2013-08-12)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
More than 10 numbers, contact manufacturer Terumo Advanced Perfusion System 1 
Manufacturer
Terumo Cardiovascular Systems Corp.
6200 Jackson Road
Ann Arbor
48103
Michigan
UNITED STATES
801763

Introduction

Recalled Products

Terumo Advanced Perfusion System 1

Reason

Terumo Cardiovascular Systems (Terumo CVS) have received three reports of a Terumo Advanced Perfusion System 1 displaying a pressure reading of '999.'  Terumo CVS has confirmed that in one of these cases, the user was unable to clear the pressure alarm and restart the pump.

Note: the system displays '999' at pressures greater than +900mmhg. The reading can indicate a true high pressure or a malfunction of the pressure transducer or cable. The operator's manual for the Terumo System 1 does not provide instruction on how to clear the '999' pressure reading.