TEC-6, TEC-6 Plus and D-TEC-6 Plus Anesthetic Vaporizers and TEC-6 Plus NAD Variet Anesthetic (2013-08-07) Aug 7, 2013 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : General Public, Healthcare Professionals, Hospitals
Department : Health Canada
Hazard classification : Type II
Brand name Product UPC Lot or serial number Companies Model or catalog number Product D. D-TEC-6 Plus Anesthetic Vaporizer Not applicable Manufacturer Datex-Ohmeda 3030 Ohmeda Drive, P.O. Box 7550, Madison 53707-7550 Wisconsin UNITED STATES 1107-9206-000 B. TEC-6 Plus Anesthetic Vaporizer Not applicable Manufacturer Datex-Ohmeda 3030 Ohmeda Drive, P.O. Box 7550, Madison 53707-7550 Wisconsin UNITED STATES 1107-9612-000 C. TEC-6 Plus - NAD Variet Anesthetic Not applicable Manufacturer Datex-Ohmeda 3030 Ohmeda Drive, P.O. Box 7550, Madison 53707-7550 Wisconsin UNITED STATES 1107-9701-000 A. TEC 6 Anesthetic Vaporizer Not applicable Manufacturer Datex-Ohmeda 3030 Ohmeda Drive, P.O. Box 7550, Madison 53707-7550 Wisconsin UNITED STATES 1107-9001-000
Recalled Products
A. TEC 6 Anesthetic Vaporizer
B. TEC-6 Plus Anesthetic Vaporizer
C. TEC-6 Plus - NAD Variet Anesthetic
D. D-TEC-6 Plus Anesthetic Vaporizer
The user manual pre-operative check, low pressure leak test may not detect the full range of leaks from seal wear degradation in the vaporizers. The integrity of the vaporizer seal may be affected in this product. If the leak is not detected, it will result in fresh gas leaking to the atmosphere and a reduction in fresh gas delivered to the breathing system.