Health Products

TEC-6, TEC-6 Plus and D-TEC-6 Plus Anesthetic Vaporizers and TEC-6 Plus NAD Variet Anesthetic (2013-08-07)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Brand nameProduct UPCLot or serial numberCompaniesModel or catalog number
Product
D. D-TEC-6 Plus Anesthetic VaporizerNot applicable
Manufacturer
Datex-Ohmeda
3030 Ohmeda Drive, P.O. Box 7550,
Madison
53707-7550
Wisconsin
UNITED STATES
1107-9206-000
B. TEC-6 Plus Anesthetic VaporizerNot applicable
Manufacturer
Datex-Ohmeda
3030 Ohmeda Drive, P.O. Box 7550,
Madison
53707-7550
Wisconsin
UNITED STATES
1107-9612-000
C. TEC-6 Plus - NAD Variet AnestheticNot applicable
Manufacturer
Datex-Ohmeda
3030 Ohmeda Drive, P.O. Box 7550,
Madison
53707-7550
Wisconsin
UNITED STATES
1107-9701-000
A. TEC 6 Anesthetic VaporizerNot applicable
Manufacturer
Datex-Ohmeda
3030 Ohmeda Drive, P.O. Box 7550,
Madison
53707-7550
Wisconsin
UNITED STATES
1107-9001-000

Introduction

Recalled Products

A. TEC 6 Anesthetic Vaporizer

B. TEC-6 Plus Anesthetic Vaporizer

C. TEC-6 Plus - NAD Variet Anesthetic

D. D-TEC-6 Plus Anesthetic Vaporizer

Reason

The user manual pre-operative check, low pressure leak test may not detect the full range of leaks from seal wear degradation in the vaporizers. The integrity of the vaporizer seal may be affected in this product. If the leak is not detected, it will result in fresh gas leaking to the atmosphere and a reduction in fresh gas delivered to the breathing system.