Health Products

Luminos DRF and YSIO (2013-09-02)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer 
Manufacturer
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
B.   YSIO10281013
3025
Manufacturer
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
A.   Luminos DRF10094200

Introduction

Recalled products
  1. Luminos DRF
  2. YSIO

Reason

During a RAD examination with automatic exposure control, a highly unlikely malfunction can result in radiation continuing longer than required. This may also result in an overexposed image that is not of diagnostic quality, so that the examination needs to be repeated. Statistically, this malfunction only occurs once every 7000 images and only if images are taken in automatic exposure mode.