Health Products

Terumo Advanced Perfusion System 1 (2013-08-12)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer 
Manufacturer
Terumo Cardiovascular Systems Corp.
6200 Jackson Road
Ann Arbor
48103
Michigan
UNITED STATES
A. Terumo Advanced Perfusion System 1 801763

Introduction

Recalled products

A. Terumo Advanced Perfusion System 1 

Reason

Terumo Cardiovascular Systems (Terumo CVS) has become aware through internal testing that the Terumo Advanced Perfusion System 1 operator's manual does not completely describe the system's centrifugal pump response under certain conditions. Specifically, if the user is adjusting the pump speed using the Central Control Monitor (CCM) or the manual speed knob on the centrifugal control unit at the same time an event occurs (which would activate the coast response mode on the pump).