Beckman Coulter Chemistry Analyzers (2013-09-11)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
All lotsF. Beckman Coulter AU480 Chemistry Analyzer
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
AU480
All lotsG. Beckman Coulter AU5800 Chemistry Analyzer
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
AU5810
All lotsD. Beckman Coulter AU640 Chemistry Analyzer
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
AU640
All lotsE Beckman Coulter AU680 Chemistry Analyzer
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
AU680
All lotsB. Beckman Coulter AU2700 Chemistry Analyzer
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
AU2700
All lotsC. Beckman Coulter AU400 Chemistry Analyzer
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
AU400
All lotsA. Beckman Coulter AU5400 Chemistry Analyzer
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
AU5400

Introduction

Recalled products
  1. Beckman Coulter AU5400 Chemistry Analyzer
  2. Beckman Coulter AU2700 Chemistry Analyzer
  3. Beckman Coulter AU400 Chemistry Analyzer
  4. Beckman Coulter AU640 Chemistry Analyzer
  5. Beckman Coulter AU680 Chemistry Analyzer
  6. Beckman Coulter AU480 Chemistry Analyzer
  7. Beckman Coulter AU5800 Chemistry Analyzer

Reason

Beckman Coulter has confirmed that some AU clinical chemistry analyzer ISE units can experience premature ISE Ref (Reference) valve failure. Premature valve failure can be detected by the presence of air bubbles in the ISE Ref solution tubing on the ISE unit. If air bubbles are present, accuracy of the ISE test results may be affected.