Health Products

Cep 8 (D8Z2) Probe (2013-08-28)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
440023Cep 8 (D8Z2) Probe 
Manufacturer
Abbott Molecular Inc.
1300 E. Touhy Ave.
Des Plaines
60018
Illinois
UNITED STATES
06J37-018

Introduction

Recalled products

A. Cep 8 (D8Z2) Probe

Reason

It has been recently identified that some kits of the CEP 8 Spectrumgreen probe, catalogue # 06J37-018 lot 440023, may have been distributed with a vial of CEP 4 Spectrumgreen probe instead of the intended CEP 8 probe.