Brilliance CT Big Bore System-Main Unit (2013-09-03)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Serial Numbers:
Manufacturer
Philips Medical Systems (Cleveland), Inc.
595 Miner Road
Cleveland
44143
Ohio
UNITED STATES
Brilliance CT Big Bore System-Main Unit 4535 670 88051

Introduction

Recalled products

A. Brilliance CT Big Bore System-Main Unit

Reason

There are artifacts appearing on the tracker images during threshold-triggered bolus scans with protocols using a rotation time of 0.4 seconds. In those cases only, the scan might be triggered either too soon or too late, which may lead to inappropriate contrast opacification during the clinical scan. If the acquired images are not diagnostically acceptable, a rescan of the patient may be performed.