Cardiohelp-I (2013-09-04)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
90410486
Manufacturer
Maquet Cardiopulmonary AG
Kehler Str. 31
Rastatt
76437
GERMANY
A. Cardiohelp-I 70104.8012

Introduction

Recalled products

A. Cardiohelp-I 

Reason

In certain rare instances the Cardiohelp-I device may display an error message with audible alarm indicating "battery defective". The alarm will occur while booting or when the main power supply is unplugged from the main power source. This is due to a defective capacitor within the battery manager board. Further, when alarming the HMI (Human Machine Interface) will gradually become black due to the HMI back light shutting down. The defective capacitor behavior leads to the shutdown of the battery booster that boosts the voltage of the batteries from 12v to 24v. This in turn leads to the state that the voltage is insufficient to ignite the backlight of the display. Blood flow is not affected during the shutdown time of the HMI, therefore there is no risk of circulatory impairment.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime