Health Products

Metasul Femoral Head and Sidus Stem-Free Shoulder Humeral Head (2013-09-05)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 serial numbers, contact manufacturer 
Manufacturer
Zimmer GmbH
Sulzer Allee 8,
Winterthur
8404
SWITZERLAND
B.    Sidus Stem-Free Shoulder Humeral HeadMore than 10 serial numbers, contact manufacturer
2607439
Manufacturer
Zimmer GmbH
Sulzer Allee 8,
Winterthur
8404
SWITZERLAND
A.    Metasul Femoral Head00-8770-040-01

Introduction

Recalled products
  1. Metasul Femoral Head
  2. Sidus Stem-Free Shoulder Humeral Head

Reason

Zimmer is initiating a recall of a subset of variously highly polished implants packaged in low density polyethylene (LDPE) bag because Zimmer has received complaints that the LDPE bag containing the implant adheres to the highly polished implant surface. The level of adhesion can vary, resulting in material transfer to the implants that can range from residue to pieces of film from the LDPE bag.