Health Products

Hill Rom 1000, Advanta 2, and Careassist P1170 Hospital Beds (2013-03-28)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Not applicable
Manufacturer
Hill-Rom Company, Inc.
1069 State Route 46 East
Batesville
47006-9167
Indiana
UNITED STATES
D. Careassist Hospital Bed Model P1170P1170
Not applicable
Manufacturer
Hill-Rom Company, Inc.
1069 State Route 46 East
Batesville
47006-9167
Indiana
UNITED STATES
B. Advanta 2 (P1190)P1190
Not applicable
Manufacturer
Hill-Rom Company, Inc.
1069 State Route 46 East
Batesville
47006-9167
Indiana
UNITED STATES
C. Advanta 2 (P1190)CS158
Not applicable
Manufacturer
Hill-Rom Company, Inc.
1069 State Route 46 East
Batesville
47006-9167
Indiana
UNITED STATES
A. Hill Rom 1000 Beds (P1160)P1160

Introduction

Recalled Products

A. Hill Rom 1000 Beds (P1160)
B. Advanta 2 (P1190)
C. Advanta 2 (P1190)
D. Careassist Hospital Bed Model P1170

Reason

The manufacturer of the power control printed circuit board (PCB) changed the control relays to another manufactures relay and that due to some switch timing issues the electrical current traveling through the relays became large enough that the relay contacts were welded together. Once the relay contacts for a function became welded, that function ceases to operate. The welding of the contacts seems to happen more when more than one function is operated at the same time.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime