B. 7 mL BD Epilor Plastic Luer-Lok, Loss of Resistance (LOR) Syringe
405198
More than 10 numbers, contact manufacturer
Manufacturer
Becton Dickinson and Company
1 Becton Drive
Franklin Lakes
07417
New Jersey
UNITED STATES
A. 7 mL BD Epilor Plastic Luer Lok, Loss of Resistance (LOR) Syringe (Component of the Durasafe Combined Spinal and Epidural Needle Set)
406078
Introduction
Recalled products
7 mL BD Epilor Plastic Luer Lok, Loss of Resistance (LOR) Syringe (Component of the Durasafe Combined Spinal and Epidural Needle Set)
7 mL BD Epilor Plastic Luer-Lok, Loss of Resistance (LOR) Syringe
Reason
BD has received several reports indicating that the plunger of the 7 mL BD Epilor plastic BD luer-lok loss of resistance syringe (LOR) may stall or stick when travelling within the syringe. This may prevent the user from detecting a loss of resistance upon entering the epidural space. If this occurs, the user may continue to advance the syringe and needle through the epidural space and the dura into the spinal canal/intrathecal space resulting in a spinal anesthesia instead of an epidural anesthesia. There have been no reports of harm to patients.
Date the event was first identified: 06/19/2013.
BD has received complaints from customers indicating that product has exhibited higher than normal resistance in the 7mL Epilor syringe. There have been a total of 7 complaints for related events in the last 3 fiscal years.