Health Products

Navios Lasers (2013-08-22)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameModel or catalog numberCompanies
More than 10 numbers, contact manufacturerB. Navios 10 Colors / 3 LasersA52103
More than 10 numbers, contact manufacturerA. Navios 8 Colors / 2 LasersA52102
Manufacturer
Beckman Coulter
Inc.
250 S. Kraemer Blvd.
92821
UNITED STATES

Introduction

Recalled products
  1. Navios 8 Colors / 2 Lasers
  2. Navios 10 Colors / 3 Lasers

Reason

Beckman Coulter has determined that the Navios Flow Cytometer instructions for use (IFU) states that when a drip chamber level warning occurs, the instrument will acquire for at least 5 minutes longer after the drip chamber level warning message is displayed. However, the system may only acquire for as little as 3 minutes after the drip chamber level warning and will provide drip chamber level error. At that time, the instrument will abort acquisition and does not save the listmode file, and data are lost.