Progeny Vantage Panoramic X-Ray System with Cephalometric (2013-09-27)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
Serial Number: ZN44018
Manufacturer
Midmark Corporation
675 Heathrow Drive
Lincolnshire
60069
Illinois
UNITED STATES
A. Progeny Vantage Panoramic X-Ray System with Cephalometric V5000C

Introduction

Recalled products

A. Progeny Vantage Panoramic X-Ray System with Cephalometric 

Reason

This field action is being initiated to correct a recently identified real time controller (RTC)firmware error found in firmware versions 3.0, 3.1, 3.2, and 3.3. This error may cause the column of the Vantage Panoramic X-Ray unit to continue moving vertically despite the release of the movement control button (i.e. "failure to stop"). If this occurs while a patient is positioned under the column, there is the potential for injury, such as bruising, sprains, or resulting from quick movements to avoid the moving column.