Health Products

Thunderbeat Instrument Grips (Pistol and Inline) and Thunderbeat Front-Actuated Grip (2013-10-02)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer
Manufacturer
Olympus Medical Systems Corp.
2951 Ishikawa-cho, Hachioji-shi,
Tokyo-to
192-8507
JAPAN
B. Thunderbeat Instrument - Inline GripTB-0510IC
More than 10 numbers, contact manufacturer
Manufacturer
Olympus Medical Systems Corp.
2951 Ishikawa-cho, Hachioji-shi,
Tokyo-to
192-8507
JAPAN
C. Thunderbeat Front-Actuated GripTB-0535FC
More than 10 numbers, contact manufacturer 
Manufacturer
Olympus Medical Systems Corp.
2951 Ishikawa-cho, Hachioji-shi,
Tokyo-to
192-8507
JAPAN
A. Thunderbeat Instrument - Pistol GripTB-0535PC

Introduction

Recalled products
  1. Thunderbeat Instrument - Pistol Grip
  2. Thunderbeat Instrument - Inline Grip
  3. Thunderbeat Front-Actuated Grip

Reason

Olympus has received customer complaints on the deterioration of the thunderbeat probe during surgical procedures. These complaints involve a combination of the following descriptions - the generation of audible and visual generator alarms, the deformation of the components of the thunderbeat probe tip or jaw or the detaching of a fragment from the thunderbeat probe tip or jaw. Olympus has investigated these claims and the manufacturer has determined that improvements to the labeling for surgical technique, including recommendations for best practices in cutting and dissection techniques during laparoscopic hysterectomy, are appropriate to reduce/eliminate these complaints associated with device use.