Trinica ALP System Case Assembly (2013-10-10)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
P061255
Manufacturer
Zimmer Spine, Inc.
7375 Bush Lake Road
Minneapolis
55439
Minnesota
UNITED STATES
A. Trinica ALP System Case Assembly 07.01058.001

Introduction

Recalled products

A. Trinica ALP System Case Assembly 

Reason

Zimmer is initiating a device correction for the Trinica anterior lumbar plate (ALP) system case assembly. The ratchet handles, Part Number 07.00438.001, may not be effectively sterilized if left in their designated location(s) in the instrument tray. This is due to the current bracketing in the tray which may interfere with complete steam penetration into the cannulated handle. To assure effective sterilization, the ratchet handles must be sterilized external to the instrument tray until newly designed replacement trays are available.