Two lots of F25ONRe Dialyzer were recalled in February, 2011 (Incident ID 61779) due to increased complaints associated with internal blood leaks. Quality trending revealed that lot 10HU02012 and 10LU02016 demonstrated a higher blood leak rate than for other lots. Following a FDA inspection, a decision was made by the legal manufacturer to place additional 20 lots as they were manufactured under the same production process and could not be definitively segregated from the recalled lots.