Health Products

Port-a-Cath II Implantable Venous Access System (2013-09-06)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
1971572
Manufacturer
Smiths Medical ASD, Inc.
1265 Grey Fox Road
St. Paul
55112
Minnesota
UNITED STATES
Port-a-Cath II Implantable Venous Access System 21-8052-24

Introduction

Recalled Products

Port-a-Cath II Implantable Venous Access System 

Reason

Since August 2009 Smiths Medical has received a total of 3 complaints regarding the introducer sheath and the introducer opening being too small for the width of the catheter experienced with Port-a-cath and Port-a-cath II implantable venous and arterial access systems and introducer sets. There have been no reports of patient injury or death associated with this issue. Investigation of this issue has identified the root cause to be related to the incorrect size  introducer being used in several lots of finished goods. An in-depth investigation concluded that the incorrect introducer size was limited to 1 component part (from St. Jude Medical), which were put into production of 36 Smiths Medical finished goods lot numbers.
Get told about the next health product recall.