Health Products

Artis Zee Biplane Systems Running Artis VC1X SW (2013-09-12)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
153146
Manufacturer
Siemens AG
Wittelsbacherplatz 2
Muenchen
80333
GERMANY
Artis Zee Biplane Systems Running Artis VC1X SW 10094141

Introduction

Recalled Products

Artis Zee Biplane Systems Running Artis VC1X SW 

Reason

If a network problem arises that puts the system into the operating mode "bypass fluoro" while the function "block radiation" is activated, this function cannot be deactivated again by once more pressing the "block radiation" key on the TSC (Touch Screen Control). Even in positions which are different from the patient transfer position and the park position, the system then displays the message "XRay manually disabled."