Some vials of Dimension Loci Cardiac Troponin I calibrator under-recover Troponin I, resulting in over-recovery of QC and patient samples following calibration. Testing on patient samples demonstrated an average upward shift of 24% (Range 1%-33%) when compared to an unaffected lot.