Health Products

Dimension EXL System-(Troponin) TNI Assay (2013-09-12)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
3DD007
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
Dimension EXL System-(Troponin) TNI Assay 10464336

Introduction

Recalled Products

Dimension EXL System-(Troponin) TNI Assay 

Reason

Some vials of Dimension Loci Cardiac Troponin I calibrator under-recover Troponin I, resulting in over-recovery of QC and patient samples following calibration. Testing on patient samples demonstrated an average upward shift of 24% (Range 1%-33%) when compared to an unaffected lot.