Health Products

ABX Pentra Calcium CP Reagent (2013-09-12)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
014664503
Manufacturer
Horiba ABX, Parc Euromedecine
rue Du Caducee - BP 7290, Cedex 4,
Montpellier
34184
FRANCE
ABX Pentra Calcium CP Reagent A11A01633

Introduction

Recalled Products

ABX Pentra Calcium CP Reagent

Reason

During QC testing, the other equipment manufacturer (OEM) discovered precipitates in some of the R2 reagent casettes of the ABX Pentra Calcium CP (A11A01633) from the lot #: 014664503. The initial functional tests, conducted by the OEM, indicate there is currently no effect on patient results. However, in a previous lot, the OEM did identify aberrant results related to the presence of these precipitates. Therefore, with time, there may be an increase in the quantity of precipitates and by extension a risk of incorrect patient results being reported.