Recalled Products
ABX Pentra Calcium CP Reagent
During QC testing, the other equipment manufacturer (OEM) discovered precipitates in some of the R2 reagent casettes of the ABX Pentra Calcium CP (A11A01633) from the lot #: 014664503. The initial functional tests, conducted by the OEM, indicate there is currently no effect on patient results. However, in a previous lot, the OEM did identify aberrant results related to the presence of these precipitates. Therefore, with time, there may be an increase in the quantity of precipitates and by extension a risk of incorrect patient results being reported.