Rapidpoint and Rapidlab Systems (2013-09-24)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
All serial numbersF. Rapidlab 1245 System - Analyzer
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave.
Tarrytown
10591
New York
UNITED STATES
1245
All serial numbersG. Rapidpoint 500 System - Main Unit
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave.
Tarrytown
10591
New York
UNITED STATES
10492730
All serial numbersD. Rapidlab 1240 System - Analyzer
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave.
Tarrytown
10591
New York
UNITED STATES
1240
All serial numbersE. Rapidlab 1265 System - Analyzer
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave.
Tarrytown
10591
New York
UNITED STATES
1265
All serial numbersB. Rapidpoint 405 System - Analyzer
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave.
Tarrytown
10591
New York
UNITED STATES
05320281
All serial numbersC. Rapidlab 1260 System - Analyzer
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave.
Tarrytown
10591
New York
UNITED STATES
1260
All serial numbersA. Rapidpoint 400 System
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave.
Tarrytown
10591
New York
UNITED STATES
106444

Introduction

Recalled Products

A. Rapidpoint 400 System
B. Rapidpoint 405 System - Analyzer
C. Rapidlab 1260 System - Analyzer
D. Rapidlab 1240 System - Analyzer
E. Rapidlab 1265 System - Analyzer
F. Rapidlab 1245 System - Analyzer
G. Rapidpoint 500 System - Main Unit

Reason

Users can inadvertently enter a wrong patient ID on the demographics screen, which ultimately could lead to a misidentification of the sample. A patient sample mismatch could be caused by:
  1. Inadvertently accepting the pre-selected patient IDon the demographics screen 
  2. Overwriting sample demographics by incorrect operation of the patient list button 
  3. Scanning the wrong barcode ID
  4. Manually typing in the wrong ID


In each of these cases, failure to check the correct demographics screen before moving on to the results screen will assign the wrong demographics to the sample being tested.