Health Products

Spectrum Infusion Pump with Master Drug Library (2013-09-24)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
997333
Manufacturer
Baxter Healthcare Corp.
711 Park Ave.
Medina
14103
New York
UNITED STATES
Spectrum Infusion Pump with Master Drug Library35700

Introduction

Recalled Products

Spectrum Infusion Pump with Master Drug Library 

Reason

Baxter Corporation is recalling one (1) Sigma Spectrum Infusion Pump with Master Drug Library that was shipped with a component that does not meet standards for withstanding an electrostatic discharge in excess of 15 kilovolts. Such exposure could cause the pump to stop unexpectedly with no audible alarm or visual display of the alarm status.