A. Advia 1200 Chemistry System - Chemistry Calibrator (2013-09-30)
Manufacturer
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave.
Tarrytown
10591
New York
UNITED STATES
09784096
Introduction
Recalled Products
A. Advia 1200 Chemistry System - Chemistry Calibrator
B. Advia Chemistry System - Chemistry Calibrators
Reason
Quality control (QC) is shifting higher impacting Direct Bilirubin (DBIL_2) with Siemens Chemistry Calibrator lots 090445B and 090445C from previous lots 090445 and 090445A. Observed QC shifts are approx. 7-9% higher with calibrator lots 090445B and 090445C. An urgent field safety notice has been issued by the manufacturer to provide affected customers with reassigned system specific values for the DBIL_2 SSVS for Chemistry Calibrator lot numbers 090445B and 090445C. Manufacturer has determined that the DBIL_2 ssvs currently assigned to these lots may cause an average shift of plus 7-9% in quality control and sample recovery at a level of 3 mg/dL. The urgent field safety notice provides reassigned values for the DBIL_2 ssvs for Chemistry Calibrator lot numbers 090445B and 090445C.