Health Products

Sapphire Multi-Therapy Infusion Pump (2013-10-08)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All serial numbers
Manufacturer
Q Core Medical Ltd.
29 Yad Haruzim St., P.O. Box 8639,
Netanya
4250529
ISRAEL
Sapphire Multi-Therapy Infusion Pump17000-028-0020

Introduction

Recalled Products

Sapphire Multi-Therapy Infusion Pump

Reason

Under certain cumulative conditions the time spent in each of the steps may deviate from the programmed time for the step. The infusion time will therefore be longer or shorter than the programmed time and the amount delivered at each step will also differ from the programmed amounts, but the total volume of the infusion will be the same as programmed.