B. Dimension Vista System - Dimension Vista 500 Analyzer
1000034648
All lots
Manufacturer
Siemens Healthcare Diagnostics Inc.
500 GBC Drive, Mailstop 514, PO Box 6101
Newark
19714-6101
Delaware
UNITED STATES
A. Dimension Vista System - Dimension Vista 1500 Analyzer
1000033793
Introduction
Recalled Products
A. Dimension Vista System - Dimension Vista 1500 Analyzer
B. Dimension Vista System - Dimension Vista 500 Analyzer
Reason
Siemens Healthcare Diagnostics has confirmed that under a rare set of conditions when utilizing the routine inventory screen to enter a user defined method (Empty) flex the system may assign the user defined method flex to a different flex that is currently in inventory on the system, and then use the incorrect flex cartridge to process the user defined method. Investigation of the issue has determined that this error will only occur if the assign empty reagent cartridge window has been left open and the flex inventory on the system changes.