A. Trabecular Reverse Shoulder Glenosphere
B. Trabecular Reverse Shoulder Base Plates
Reason
Zimmer is initiating a product recall for the listed Zimmer trabecular metal reverse shoulder system glenosphere and base plate item numbers because it has received 90 complaints (complaint rate of 0.27%) related to the glenosphere disassociating from the base plate or difficulty assembling the glenosphere to the base plate. Based on the investigation of these complaints, Zimmer determined that additional instructions specific to the assembly of the glenosphere to the base plate were necessary in order to emphasize important techniques.