Medtronic Xomed Instrument Trays (2013-07-24)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer
Manufacturer
Medtronic Xomed Inc.
6743 Southpoint Drive North
Jacksonville
32216
Florida
UNITED STATES
A. Medtronic Xomed Instrument Trays More than 10 numbers, contact manufacturer

Introduction

Recalled products

A. Medtronic Xomed Instrument Trays 

Reason

During the validation of a new instrument, testing revealed that use of a gravity steam cycle with one of the listed instrument trays did not produce the sterility assurance level (SAL) of 10-6 recommended by industry standard (ANSI / AAMI ST77:2006). A subsequent investigation of all Medtronic Xomed instrument trays identified additional trays as having certain locations within the tray that would not meet the recommended sterilization level of SAL10-6 during gravity steam sterilization cycles. The trays identified through the testing are the subject trays involved in this field action.

Instruments that are sterilized in the "worst-case" locations of the identified trays have a higher potential to be inadequately sterilized. An instrument that is inadequately sterilized can become the source of cross-contamination or cross-infection when used in multiple surgical procedures.

At this time, Medtronic has no evidence in the complaint files to indicate there have been any reports of inadequate sterilization cycles (i.e. Cycle failures) or adverse events (Infections) caused by inadequate sterilization of instruments in the subject instrument trays.

Note: this issue only involves gravity steam sterilization cycles. The listed instrument trays have been demonstrated to be satisfactorily validated for prevacuum steam sterilization cycles as detailed in the instructions for use that accompany Medtronic Xomed and Medtronic Xomed instrumentation manual surgical instruments.