Health Products

Avea Ventillator (2013-09-03)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
More than 10 serial numbers, contact manufacturerA. Avea Ventillator 
Manufacturer
Carefusion
22745 Savi Ranch Parkway
Yorba Linda
92887
California
UNITED STATES
AVEA

Introduction

Recalled products

A. Avea Ventillator 

Reason

Avea ventilators may experience the under reporting of tidal volume if used in conjunction with the neonatal hotwire flow sensor, Carefusion part number 16465.  This error is due to the barometric pressure sensor compensation.  Clinically significant changes in barometric pressure are generally seen at altitudes greater than 5,000 feet above sea level (ASL).  Single patient use Varflex sensors are not affected by this corrective action and will compensate for changes to atmospheric barometric pressure.  The adult and pediatric patient setting are not affected by this issue.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime