Homechoice and Homechoice Pro Automated PD Systems (2013-10-10) Oct 10, 2013 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : General Public, Healthcare Professionals, Hospitals
Department : Health Canada
Hazard classification : Type II
Lot or serial number Companies Brand name Model or catalog number More than 1000 numbers, contact manufacturer Manufacturer Baxter Healthcare Corporation 1 Baxter Parkway Deerfield 60015 Illinois UNITED STATES B. Homechoice Pro Automated PD System 5C8310 More than 1000 numbers, contact manufacturer Manufacturer Baxter Healthcare Corporation 1 Baxter Parkway Deerfield 60015 Illinois UNITED STATES A. Homechoice Automated PD System 5C4471
Recalled products
Homechoice Automated PD System
Homechoice Pro Automated PD System
For dialysis patients with ascites using Homechoice and Homechoice Pro devices with software version 10.4, because the device will attempt to drain the patient to empty during the initial drain, there is an increased risk of hypotension and compromised hemodyamic status.