Health Products

Vitros Chemistry Products Mixing Cup Arrays (2013-10-11)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer 
Manufacturer
Ortho-Clinical Diagnostics, Inc.
100 Indigo Creek Drive
Rochester
14626
New York
UNITED STATES
A. Vitros Chemistry Products Mixing Cup Arrays 1631779

Introduction

Recalled products

A. Vitros Chemistry Products Mixing Cup Arrays 

Reason

Ortho clinical diagnostics (OCD) has become aware of an increase in EAE condition codes when using Vitros 250 or 350 chemistry systems. When customers perform an on-analyzer dilution, it is performed by the Vitros chemistry system using the Vitros mixing cup arrays and diluents. OCD's investigation has found that while performing on-analyzer dilution, some customers have reported an increase in EAE conditions codes ("Sample metering dilution error"). When an EAE condition code occurs, the processing for that patient sample is automatically cancelled by the system and the user is alerted.

OCD's investigation has determined that the increased rate of EAE condition codes is attributed to out-of-specification Vitros mixing cup arrays. OCD is working with the OEM manufacturer of the mixing cup arrays to determine the root cause of the issue.