Neuron Max 088 Long Sheath (2013-09-19)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
F31580
Manufacturer
Penumbra Inc.
1351 Harbor Bay Parkway
Alameda
94502
California
UNITED STATES
Neuron Max 088 Long SheathPNML6F088804M

Introduction

Recalled Products

Neuron Max 088 Long Sheath

Reason

Penumbra has become aware of complaints involving devices where the operator complained of damage to the tip of the Neuron Max MP catheter after removal from the packaging. Upon evaluation of the returned devices, Penumbra confirmed that the polymer at the distal tip of the device had been damaged. Further investigation determined that the sticker adhesive used to secure the packaging mandrel to the packaging card was less sticky then previous lots of stickers.