Allset Gold and SSP Unitary Low Resolution Kits (2013-08-30)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type III
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
More than 10 numbers, contact manufacturer D.    SSP Unitary - Low Resolution Class I
Manufacturer
Life Technologies Corporation (Invitrogen)
9099 North Deerbrook Trail
Brown Deer
53223
Wisconsin
UNITED STATES
78000-10
038 1249868B.     Allset Gold ABC Low Resolution Kit
Manufacturer
Life Technologies Corporation (Invitrogen)
9099 North Deerbrook Trail
Brown Deer
53223
Wisconsin
UNITED STATES
54340D
029 1247232C.    Allset Gold ABDRQ Low Resolution Kit
Manufacturer
Life Technologies Corporation (Invitrogen)
9099 North Deerbrook Trail
Brown Deer
53223
Wisconsin
UNITED STATES
54360D
022 1256853A. Allset Gold HLA Low Resolution Kits
Manufacturer
Life Technologies Corporation (Invitrogen)
9099 North Deerbrook Trail
Brown Deer
53223
Wisconsin
UNITED STATES
54310D

Introduction

Recalled Products

A. Allset Gold HLA Low Resolution Kits
B. Allset Gold ABC Low Resolution Kit
C. Allset Gold ABDRQ Low Resolution Kit
D. SSP Unitary - Low Resolution Class I

Reason

During an internal complaint investigation for problem report PR#135598, the reactivity assignment in our Human Oligotyping software (HOS) program for A32:04 allele in primer mix PM019e was determined to be incorrect. The catalog and lot numbers of the products listed above have been identified to be affected by the incorrect reactivity assignment. The kits would produce an incorrect low resolution typing or mistyping result of A*02, A*03 alleles when testing a sample which has the A*32:04 allele. The user would be unaware of the incorrect type, unless they confirm the sample typing by another method, therefore, this would be considered a mistype.