Vitros Immunodiagnostic Products Anti-HBS Calibrator and Reagent Pack
Ortho Clinical Diagnostics (OCD) is notifying customers of the potential for non-reproducible and unexpected positive results for negative Anti-HBS reagent packs, lots 2280, 2290, 2302 and 2310. These lots were distributed beginning in November 2012, and all lots have now expired.
OCD internal data indicates that the rate of occurrence is variable, the maximum observed rate of occurrence for the affected lots was 4 out of 104 negative Anti-HBS patient samples, which represents approximately 3.9% of negative samples tested. Users are being advised to review all previously reported results between more than 12 and less than 30 mIU/mL. A false positive result considers that a person is presumed to be immune to an HBV infection leaving the person susceptible to an HBV infection.