Vitros Immunodiagnostic Products Anti-HBS Calibrator and Reagent Pack (2013-10-25)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
2280
Manufacturer
Ortho-Clinical Diagnostics
Johnson & Johnson
500-100 Holmers Farm Way
High Wycombe, Buckinghamshire
HP12 4DP
UNITED KINGDOM
Vitros Immunodiagnostic Products Anti-HBS Calibrator and Reagent Pack1524693

Introduction

Recalled Products

Vitros Immunodiagnostic Products Anti-HBS Calibrator and Reagent Pack

Reason

Ortho Clinical Diagnostics (OCD) is notifying customers of the potential for non-reproducible and unexpected positive results for negative Anti-HBS reagent packs, lots 2280, 2290, 2302 and 2310. These lots were distributed beginning in November 2012, and all lots have now expired.

OCD internal data indicates that the rate of occurrence is variable, the maximum observed rate of occurrence for the affected lots was 4 out of 104 negative Anti-HBS patient samples, which represents approximately 3.9% of negative samples tested. Users are being advised to review all previously reported results between more than 12 and less than 30 mIU/mL. A false positive result considers that a person is presumed to be immune to an HBV infection leaving the person susceptible to an HBV infection.