Health Products

ABL90 Flex Analyzer (2013-10-24)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberBrand nameCompaniesModel or catalog number
N/AABL90 Flex Analyzer
Manufacturer
Radiometer Medical APS
Akandevej 21
Bronshoj
2700
DENMARK
393-090

Introduction

Recalled Products

ABL90 Flex Analyzer

Reason

If the ABL90 flex analyzer is tilted it may cause a negative bias on sodium results up to 7% (minus 10mm).  This issue has been reported when the analyzer has been installed on a roll stand and tilted (i.e. the support plate is not horizontal).