Health Products

Anaconda Body Delivery Systems (2013-10-21)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
More than 10 numbers, contact manufacturer 
Manufacturer
Vascutek Ltd.
Newmains Avenue, Inchinnan Industrial Estate
Renfrewshire, Scotland,
PA4 9RR
UNITED KINGDOM
B. Anaconda One-Lok Bifurcate BodyOLB21
All serial numbers 
Manufacturer
Vascutek Ltd.
Newmains Avenue, Inchinnan Industrial Estate
Renfrewshire, Scotland,
PA4 9RR
UNITED KINGDOM
A. Anaconda Bifurcate Body StentB19

Introduction

Recalled Products

A. Anaconda Bifurcate Body Stent
B. Anaconda One-Lok Bifurcate Body

Reason

Vascutek has received 3 complaints (with an occurrence rate of 0.1%) where the release wire of an Anaconda Body Delivery system has fractured following stent deployment. Two of these three incidents resulted in conversion to open repair. The manufacturer is recalling all Anaconda bodies. Previously implanted stents are not affected by this recall.
Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime