Health Products

Sysmex UF-100 Instrument Manager Software (2013-11-06)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
N/A
Manufacturer
Data Innovations LLC
120 Kimball Ave. Suite 100
South Burlington
05403
Vermont
UNITED STATES
Instrument Manager Software using the Sysu100I Driver that connects to the Sysmex UF-100 Instrument7.00.0001

Introduction

Recalled Products

Instrument Manager Software using the Sysu100I Driver that connects to the Sysmex UF-100 Instrument

Reason

When results of "---.-" are received from the Sysmex UF-100 instrument and if the user set the driver configuration option "Strip zeros from result" to "Yes", all versions prior to V7.00.0006 and V8.00.0005 of the SYSU100I driver convert the results to "0.00" in instrument manager. The Sysmex UF-100 will send "--.-" when the test has a sample error. All five urinalysis tests performed (RBC, WBC, Ec, Cast, and Bact) will be resulted as "0.00" in instrument manager. This could lead to misdiagnosis and mistreatment of a patient.