Medtronic Steerable Guidewire (2013-10-21) Oct 21, 2013 · Published by Health Canada · Official notice
Share
Alert type : Medical Device Recall
Category : Medical Device
Issue : Medical Devices
Audience : General Public, Healthcare Professionals, Hospitals
Department : Health Canada
Hazard classification : Type I
Lot or serial number Brand name Model or catalog number Companies More than 10 numbers, contact manufacturer D. Attain Hybrid Guide Wire GWR419478 More than 10 numbers, contact manufacturer B. Cougar LS Steerable Guidewire CGRLS190HJ More than 10 numbers, contact manufacturer C. Thunder Steerable Guidewire THNDR190J More than 10 numbers, contact manufacturer A. Cougar XT Steerable Guidewire CGRXT190HJ Manufacturer Medtronic Inc. 710 Medtronic Parkway NE Minneapolis 55432 Minnesota UNITED STATES
Recalled products
Cougar XT Steerable Guidewire
Cougar LS Steerable Guidewire
Thunder Steerable Guidewire
Attain Hybrid Guide Wire
Medtronic has identified the potential for an issue with a specific subset of lot numbers of the medtronic interventional wires and attain hybrid guidewires where the polytetrafluoroethylene (PTFE) coating could delaminate and detach from the guidewire.