I-Stat 1 Analyzer System - PT/LNR Cartridge and Controls (2013-10-28)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
C13139 to C13270A
Manufacturer
Abbott Point of Care Inc.
100 and 200 Abbott Park Rd.
Abbott Park
60064
Illinois
UNITED STATES
A. I-Stat 1 Analyzer System - PT/LNR Cartridge and Controls03P89-24

Introduction

Recalled products

A. I-Stat 1 Analyzer System - PT/LNR Cartridge and Controls

Reason

Abbott Point of Care has determined that I-stat PT/INR cartridges have the potential to exhibit incorrectly elevated results. Internal studies have demonstrated that I-stat PT/INR results are elevated by an average of approximately 10% as compared to the international reference preparation (RTF/09) in the therapeutic range of 1.8 to 3.0 INR.