A. I-Stat 1 Analyzer System - PT/LNR Cartridge and Controls
03P89-24
Introduction
Recalled products
A. I-Stat 1 Analyzer System - PT/LNR Cartridge and Controls
Reason
Abbott Point of Care has determined that I-stat PT/INR cartridges have the potential to exhibit incorrectly elevated results. Internal studies have demonstrated that I-stat PT/INR results are elevated by an average of approximately 10% as compared to the international reference preparation (RTF/09) in the therapeutic range of 1.8 to 3.0 INR.