Health Products

Altrus Thermal Tissue Fusion Handpieces (2013-11-01)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
13CHB006
Manufacturer
Conmed Corporation
525 French Road
Utica
13502
New York
UNITED STATES
A. Altrus Thermal Tissue Fusion Handpieces60-9522-001

Introduction

Recalled products

A. Altrus Thermal Tissue Fusion Handpieces

Reason

Conmed Corporation confirmed that two lots of Altrus Thermal Tissue Fusion handpiece devices were exposed to a higher than acceptable level of particulate matter during manufacture in March 2013. Device sterility and performance are unaffected by this exposure. These lots were released for distribution without following the specific documentation requirements of Conmed electrosurgery to place the affected products on non-conformance report prior to disposition. In October 2013, conmed notified the USFDA of their intent to recall Altrus Thermal Tissue Fusion handpieces catalog no. 60-9520-001, lot 13CHB007, and 60-9522-001, lot 13CHB006, due to exposure to high particle count during manufacture. Only catalog no. 60-9522-001, lot 13CHB006, was distributed in Canada.