Porte-implant Synfix-LR (2013-11-06)

· Published by Health Canada · Official notice

Details

Alert type: Medical Device Recall
Category: Medical Device
Issue: Medical Devices
Audience: General Public, Healthcare Professionals, Hospitals
Department: Health Canada
Hazard classification: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberCompaniesBrand nameModel or catalog number
All lots
Manufacturer
Synthes (Canada) Ltd.
2566 Meadowpine Blvd.
Mississauga
L5N 6P9
Ontario
CANADA
A. Synfix-LR Implant Holder 03.802.039

Introduction

Recalled products

A. Synfix-LR Implant Holder 

Reason

Complaints have been received which describe the Synfix LR Implant Holder breaking at the interface between the implant and holder. If an unretrieved device fragment remains threaded into the plate, it will not be possible to properly attach the Synfix aiming device to the implant (Plate). Proper attachment of the aiming device to the plate is required for accurate insertion of the four Synfix screws into the Synfix implant (Plate) and verebral bodies. If the tip holder should break, the potential exists for an untretrieved device fragment to be left in the Synfix-LR implant.