B. Unicel DXH 600 Coulter Cellular Analysis System
B23858
All lots
Manufacturer
Beckman Coulter Inc.
250 S. Kraemer Blvd.
Brea
92821
California
UNITED STATES
A. Unicel DXH 800 Coulter Cellular Analysis System
629029
Introduction
Recalled products
Unicel DXH 800 Coulter Cellular Analysis System
Unicel DXH 600 Coulter Cellular Analysis System
Reason
Complaints related to optical degredation have increased on the DXH 800 and DXH 600 systems. Customers have reported that they are experiencing an increase in one or more of the following: (1) Flow cell clogs; (2) ls offset errors during daily check or control analysis; (3) High coefficients of variation for Latron CP-X controls for low angle light scatter and/or axial light loss parameters; (4) System messages and R-Flagged results for NRBC, Diff and Retic during control or patient analysis, such as partial clogs (Low event rated with Diff RS) or a voltage failure (System event: D with Diff RS).